MVP-10 Quick Check Out & Set-up Guide
- The following steps may be performed prior to use to ensure the ventilator appears to be functioning correctly.
- A more detailed procedure is indicated in the manual for Clinical Engineering or Biomedical personnel.8
Quick Performance Verification
- Attach circuit and occlude wye.
- Turn flow to 12 lpm.
- Set switch to ‘Cycle’.
- Set Insp. Time to .4 and Exp. Time to .6 and verify that the ventilator cycles at a rate of 60 breaths per minute.
- Turn PEEP knob to maximum position (fully counter-clockwise) and verify that at least 15 cm of PEEP is present on the manometer.
- Turn Max. Pressure knob to maximum position (fully counter-clockwise) and verify that appropriate pressure relief (pop-off) level is present during the inspiratory phase on the manometer (Note: You may need to increase pressure relief setting). With pressure relief valve set at maximum position 65 cm of pressure should be attained.
Set-Up For Pressure Limited Ventilation
- Attach patient circuit and occlude wye.
- Turn flow to desired level.
- Set switch to ‘Cycle’.
- Set desired inspiratory time.
- Adjust expiratory time to set respiratory rate
- Adjust pressure relief (pop-off) valve if desired to an appropriate level above intended peak pressure setting (either occlude exhalation valve or turn Max Pressure knob to max setting before setting pressure relief valve)
- Adjust Max Pressure knob until desired Peak Pressure is observed on manometer.
- Adjust PEEP knob until desired PEEP is displayed during expiration on manometer.
Set-Up For Volume Limited Ventilation
- Place appropriate sized test lung on circuit (infant or pediatric depending on desired tidal volume) instead of occluding it.
- Calculate correct flow setting using the following two formulas:
- Flow Rate Multiplication Factor = 60
insp. Time (seconds)
- Flow Rate (liters per minute) = Desired Tidal Volume (ml) x multiplication factor
1000
- Ensure that Max Pressure knob is NOT pressure limiting the volume breath.
- Set pressure relief valve to appropriate safety level.